Post-elution concentration of 99mTcO4- by a single anion exchanger column. I: Feasibility of extending the useful life of column chromatographic 99mTc generator.

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine(2001)

引用 7|浏览1
暂无评分
摘要
99mTc eluate in 0.7/0.35 M acetic acid and in 1:1 v/v 0.7 M acetic acid: 0.025 M ammonium acetate (30-40 ml) was conveniently concentrated by passing through a small column of a readily available weak anion exchanger, DEAE cellulose, (200 mg, 6 x 8 mm) and subsequent elution with 3 ml normal saline, with greater than 90% recovery. The quality of pertechnetate thus obtained and its compatibility for preparing 99mTc compounds were satisfactory. This method was successfully applied to the alumina column chromatographic generator produced in-house and to a commercial 99Mo-99mTc generator after extensive clinical use. Potential for much higher concentration factors with compacted anion exchanger cartridges as well as applicability for 188W-188Re generator are discussed.