Observation of X-ray and Auger electron spectra in a 4π proportional counter for 4π(e, X)-γ coincidence measurements.

Applied Radiation and Isotopes(2014)

引用 2|浏览1
暂无评分
摘要
For the standardization of electron capture nuclides followed by γ-transitions, the 4π(e, X)–γ coincidence counting method is mainly adopted. To examine how the spectra of the (e, X) channel change with gas nature and pressure, pulse-height spectrum measurements of signals from a 4π pressurized proportional counter were carried out under various conditions for 54Mn and 88Y sources together with 4π(e, X)–γ coincidence counting. The spectra were measured for each half of the counter and for the combined 4π signal. The nature of counting gas (P-10 or methane) and its pressure was reflected in the X-ray peak intensity, and the Auger electron spectra were influenced significantly by the absorption in a thin VYNS film and the source conditions.
更多
查看译文
关键词
Pulse-height spectrum,4π proportional counter,Electron capture,X-rays,Auger electrons,4π(e, X)–γ coincidence counting