Force-controlled dynamic wear testing of total ankle replacements (vol 12, pg 332, 2015)
Acta Biomaterialia(2015)
摘要
Currently, our knowledge of wear performance in total ankle replacements is limited. The aim of this study is to develop a scenario for force-controlled testing and wear testing of total ankle replacements. A force-controlled wear test was developed: based on cadaver measurements, the passive stabilization (ligaments and soft tissue) of the ankle joint was characterized and a restraint model for ankle stabilization was developed. Kinematics and kinetics acting at the replaced ankle joint were defined based on literature data and gait analysis. Afterwards, force-controlled wear testing was carried out on a mobile, three-component, total ankle replacement design. Wear was assessed gravimetrically and wear particles were analyzed. Wear testing resulted in a mean wear rate of 18.2±1.4mm3/106 cycles. Wear particles showed a mean size of 0.23μm with an aspect ratio of 1.61±0.96 and a roundness of 0.62±0.14. Wear testing of total ankle replacement shows that a relevant wear mass is generated with wear particles in a biologically relevant size range. The developed wear test provides a basis for future wear testing of total ankle replacements.
更多查看译文
关键词
Wear,Ankle replacement,Ankle arthroplasty,Wear particles,Wear simulation
AI 理解论文