Comparison of the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 test v1.0 with v2.0 in HIV-1 viral load quantification.

The Kaohsiung journal of medical sciences(2015)

引用 3|浏览3
暂无评分
摘要
Roche modified the COBAS AmpliPrep/COBAS TaqMan human immunodeficiency virus type 1 (HIV-1) test version 1.0 (CAP/CTM v1.0), resulting in the COBAS AmpliPrep/COBAS TaqMan HIV-1 test version 2.0 (CAP/CTM v2.0). The aim of this study was to evaluate the performance of the CAP/CTM v2.0 and to compare this performance with that of the CAP/CTM v1.0. The study was conducted in a small local study group in Kaohsiung Medical University Hospital, Kaohsiung, Taiwan. A total of 86 plasma samples from HIV-1-seropositive patients were tested using the two assays. The correlation and concordance of results between the two assays were calculated. The CAP/CTM v2.0 generated higher values than did the CAP/CTM v1.0, and five samples (5.8%) yielded a difference of > 1 log10 copies/mL. In addition, our data show that CAP/CTM v1.0 and CAP/CTM v2.0 yielded relatively consistent values for 23 samples with low viral loads (< 200 copies/mL). Furthermore, when viral loads were in a medium range (2-5 log10 copies/mL), the results of the two assays were more compatible. This study shows a good correlation between CAP/CTM v1.0 and v2.0 in HIV-1 viral load measurement. Further attention must be paid to those cases in which measured viral loads present larger differences between the two assays.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要