FIELD CORRECTION FACTORS FOR PERSONAL NEUTRON DOSEMETERS.

RADIATION PROTECTION DOSIMETRY(2016)

引用 2|浏览11
暂无评分
摘要
A field-dependent correction factor can be obtained by comparing the readings of two albedo neutron dosemeters fixed in opposite directions on a polyethylene sphere to the H*(10) reading as determined with a thermal neutron detector in the centre of the same sphere. The work shows that the field calibration technique as used for albedo neutron dosemeters can be generalised for all kind of dosemeters, since H*(10) is a conservative estimate of the sum of the personal dose equivalents H-p(10) in two opposite directions. This result is drawn from reference values as determined by spectrometers within the EVIDOS project at workplace of nuclear installations in Europe. More accurate field-dependent correction factors can be achieved by the analysis of several personal dosimeters on a phantom, but reliable angular responses of these dosemeters need to be taken into account.