A Bidirectional Neural Interface Circuit With Active Stimulation Artifact Cancellation and Cross-Channel Common-Mode Noise Suppression

Solid-State Circuits, IEEE Journal of(2016)

引用 85|浏览11
暂无评分
摘要
This work presents a bidirectional neural interface circuit that enables simultaneous recording and stimulation with a stimulation artifact cancellation circuit. The system employs a common average referencing (CAR) front-end circuit to suppress cross-channel environmental noise to further facilitate use in clinical environment. This paper also introduces a new range-adapting (RA) SAR ADC to lower the system power consumption. A prototype is fabricated in 0.18 μm CMOS and characterized and tested in vivo in an epileptic rat model. The prototype attenuates stimulation artifacts by up to 42 dB and suppresses cross-channel noise by up to 39.8 dB. The measured power consumption per channel is 330 nW, while the area per channel is $0.17,hbox{mm}^{2}$.
更多
查看译文
关键词
Closed-loop stimulation,SAR ADC,low power,neural recording,noise cancellation