Inter-Rater and Intra-Rater Repeatability and Reliability of EOS 3-Dimensional Imaging Analysis Software.

The Journal of Arthroplasty(2016)

引用 17|浏览7
暂无评分
摘要
Background: Quantifying ideal component position for the acetabulum and stem during total hip arthroplasty (THA) has been described by many methods. A new imaging method using low-dose digital stereoradiography, the EOS imaging system, is a biplanar low-dose X-ray system that allows for 3-dimensional modeling of lower limbs and semiautomated measurement of pelvic parameters and implant alignment. Methods: Twenty-five patients who underwent primary THA by a single surgeon between October 2014 and December 2014 were retrospectively selected. Only patients with unilateral THA without associated spine pathologies were included, totaling 16 right hips and 9 left hips. There were 8 men and 17 women in the cohort, with a mean age of 67 years (range, 53-82). Three individuals performed measurements of pelvic parameters and implant alignment on 3 separate occasions. An interclass correlation of >0.75 was accepted as evidence of excellent agreement and a confirmation of measurement reliability. Results: Before reviewing patient radiographs, 4 pelvic phantom models were analyzed using the EOS 3-dimensional software to verify accuracy. All anatomic and implant measurements performed by the 3 independent reviewers showed interobserver and intraobserver agreement with interclass correlation >0.75. Conclusion: Three-dimensional modeling of hip implants with the EOS imaging system is a reasonable option for the evaluation of component position after THA. (C) 2016 Elsevier Inc. All rights reserved.
更多
查看译文
关键词
total hip arthroplasty,radiographic analysis,EOS imaging
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要