LC-MS/MS multiplexed assay for the quantitation of a therapeutic protein BMS-986089 and the target protein Myostatin.

BIOANALYSIS(2016)

引用 23|浏览8
暂无评分
摘要
Background: Therapeutic protein discovery study highlights the need for the development of quantitative bioanalytical methods for determining the levels of both the therapeutic protein and the target protein, as well. Results: For the quantitation of BMS-986089, both accuracy (99-103%) and precision (2.4-12%) were obtained for the analysis of the surrogate peptide (ITYGGNSPVQEFTVPGR), in addition to the accuracy (100-108%) and precision (0.7-18%) that were obtained for the analysis of the surrogate peptide (VVSVLTVLHQDWLNGK). For Myostatin, accuracy (94-103%) and precision (2.4-14.9%) were obtained for the analysis of the surrogate peptide (IPAMVVDR). Conclusion: The developed method was applied to the analysis of samples following dosing of BMS-986089 to mice. This method highlights the potential of LC-MS/MS-based methods to eventually assess in vivo drug-target engagement.