A new method to standardize CBCT for quantitative evaluation of alveolar ridge preservation in the mandible: a case report and review of the literature.

Regenerative biomaterials(2015)

引用 5|浏览19
暂无评分
摘要
Cone-beam computerized tomography (CBCT) is an effective technique for assessment of changes to the alveolar ridge (AR). However, its accuracy and reliability could be improved by standardization of imaging positions to remain unchanged during measurements. In this study, an alveolar ridge preservation procedure was performed on a left third molar (38) socket by filling it with a radiotransparent synthetic bone graft, mineralized collagen (MC). Photographic, X-ray and CBCT images were captured before and 3, 6 and 12 months after surgery. A new method was developed to standardize CBCT for quantitative evaluation. Obtained CBCT images showed good comparability. The post-extraction alveolar width and height were both over 95% of the original values, but some resorption of the lingual bone wall (>50%) and inter-crestal bone (>30%). It is concluded that an effective positional standardization method was developed for CBCT assessment of AR dimensional changes in the posterior mandible. The use of MC in combination with a collagen membrane improved dimensional preservation of the AR.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper