Calibration Procedures For A 10.6 Mu M Laser Calorimeter

FIFTH CONFERENCE ON OPTICS (ROMOPTO '97), PTS 1 AND 2(1998)

引用 1|浏览5
暂无评分
摘要
The paper presents the theoretical background of laser calorimetry as a method to investigate the absorptance of optical components, according to the recently adopted ISO Standard 11 551. An experimental set-up for optical absorption measurements, including a cw CO2 laser and a calorimetric chamber, is presented. The calibration of the laser calorimetric set-up is also described. The calibration involved several steps: laser power calibration, evaluation of chamber sensitivity, electrical calibration using electrically heated samples and optical calibration utilizing 100 % absorptance samples. There were measured the conversion factors for the laser power and for the sample temperature; there were calculated over-all correction factors for the different substrates and they could be placed within +/-4 % around their mean value.
更多
查看译文
关键词
laser optical component, absorptance, ISO standard, laser calorimetry, calibration