谷歌浏览器插件
订阅小程序
在清言上使用

Robotically Assisted Velocity-Sensitive Triggered Focused Ultrasound Surgery

AIP conference proceedings(2012)

引用 0|浏览12
暂无评分
摘要
Magnetic Resonance (MR) guided Focused Ultrasound Surgery (FUS) of abdominal organs is challenging due to breathing motion and limited patient access in the MR environment. In this work, an experimental robotically assisted FUS setup was combined with a MR-based navigator technique to realize motion-compensated sonications and online temperature imaging. Experiments were carried out in a static phantom, during periodic manual motion of the phantom without triggering, and with triggering to evaluate the triggering method. In contrast to the non-triggered sonication, the results of the triggered sonication show a confined symmetric temperature distribution. In conclusion, the velocity sensitive navigator can be employed for triggered FUS to compensate for periodic motion. Combined with the robotic FUS setup, flexible treatment of abdominal targets might be realized.
更多
查看译文
关键词
Medical Robotics,Magnetic Resonance Imaging,MRgFUS,HIFU
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要