Simultaneous laser vibrometry on multiple surfaces with a single beam system using range-resolved interferometry

Proceedings of SPIE(2015)

引用 9|浏览1
暂无评分
摘要
A novel range-resolved interferometric signal processing technique that uses sinusoidal optical frequency modulation is applied to multi-surface vibrometry, demonstrating simultaneous optical measurements of vibrations on two surfaces using a single, collimated laser beam, with a minimum permissible distance of 3.5 cm between surfaces. The current system, using a cost-effective laser diode and a fibre-coupled, downlead insensitive setup, allows an interferometric fringe rate of up to 180 kHz to be resolved with typical displacement noise levels of 8 pm Hz(-0.5). In this paper, the system is applied to vibrometry measurements of a table-top cryostat, with concurrent measurements of the optical widow and the sample holder inside. This allows the separation of common-mode vibrations of the whole cryostat from differential vibrations between the window and the sample holder.
更多
查看译文
关键词
Interferometry,Vibrometry,Displacement Sensing,Range-Resolved Interferometric Signal Processing,LDV,Multi-Surface Vibrometry,Cryogenics,Low temperature