Evaluation of Potentially Influential Factors for Positron Emission Tomography Image Quality in Liver Signal-to-noise Ratio Utilizing a Delivery 18F-fluoro-2-deoxy-D-glucose Study on Bi4Ge3O12-Positron Emission Tomography/Computed Tomography

Hideo Onishi,Yoshiharu Kangai, Shinichi Yanagimoto, Hiroaki Mimura, Hiroyasu Sanai,Kunihiko Yoshikawa, Atsushi Takigawa

Japanese Journal of Radiological Technology(2014)

引用 23|浏览2
暂无评分
摘要
The purpose of this study was to evaluate the relationship of body habitus, blood glucose level and injected dose, respectively, with BGO (Bi4Ge3O12) positron emission tomography (PET) image quality using commercially available 2-deoxy-2-[18F] fluoro-D-glucose (FDG). We also evaluated the relationship between PET image quality and acquisition time for each weight group. Method: One hundred twenty-five patients (66 male, 59 female) were enrolled in the study. We adopted liver signal-to-noise ratio (liver SNR) as an image quality index, derived from the region of interest (ROI) placed on the axial image of the liver. Results: The correlation coefficient between liver SNR and dose per weight was 0.502. The liver SNR indicated a negative relationship with body weight, body mass index (BMI) and cross sectional area of the patient’s body, with the correlation coefficients of −0.594, −0.479 and −0.522, respectively. For all weight groups, an extended acquisition time of at least 60 s/bed was necessary to improve liver SNR. Conclusion: The findings of this study are potentially of use for designing imaging protocols for the BGO-PET/CT system when using commercially available FDG. It is easy to obtain good image quality for patients of low to average body size with the standard injection dose. However, large patients should be injected, wherever possible, with an FDG dose of up to 5 MBq/kg. The acquisition time in overweight and obese patients should be as longer as possible than in standard weight patients.
更多
查看译文
关键词
Image Quality,Imaging
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要