DC and AC biasing of a transition edge sensor microcalorimeter
AIP CONFERENCE PROCEEDINGS(2002)
摘要
We are developing AC-biased transition edge sensor (TES) microcalorimeters for use in large arrays with frequency-domain multiplexing. Using DC bias, we have achieved a resolution of 17 eV FWHM at 2.6 keV with a decay time of 90 As and an effective detector diameter of 300 Am. We have successfully measured thermal pulses with a TES microcalorimeter operated with an AC bias. We present here preliminary results from a single pixel detector operated under DC and AC bias conditions.