Primary standard for liquid flow rates between 30 and 1500 nl/min based on volume expansion.

BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK(2015)

引用 14|浏览11
暂无评分
摘要
An increasing number of microfluidic systems operate at flow rates below 1 mu l/min. Applications include (implanted) micropumps for drug delivery, liquid chromatography, and microreactors. For the applications where the absolute accuracy is important, a proper calibration is required. However, with standard calibration facilities, flow rate calibrations below similar to 1 mu l/min are not feasible because of a too large calibration uncertainty. In the current research, a traceable flow rate using a certain temperature increase rate is proposed. When the fluid properties, starting mass, and temperature increase rate are known, this principle yields a direct link to SI units, which makes it a primary standard. In this article, it will be shown that this principle enables flow rate uncertainties in the order of 2-3% for flow rates from 30 to 1500 nl/min.
更多
查看译文
关键词
calibration,comparison,implanted infusion pumps,micro and nano liquid flow rates,primary standard,uncertainty,validation,volumetric expansion