Off-Line, Real-Time, Ft-Ir Analysis Of A Process Step In Imipenem Production

Jhansi R Boaz, Scott M Thomas,Steven M Meyerhoffer,Steven J Staskiewicz, Joseph E Lynch,Richard S Egan, Dean Ellison

OPTICALLY BASED METHODS FOR PROCESS ANALYSIS(1992)

引用 0|浏览17
暂无评分
摘要
We have developed an FT-IR method, using a Spectra-Tech Monit-IR 400 systems, to monitor off-line the completion of a reaction in real-time. The reaction is moisture-sensitive and analysis by more conventional methods (normal-phase HPLC) is difficult to reproduce. The FT-IR method is based on the shift of a diazo band when a conjugated beta-diketone is transformed into a silyl enol ether during the reaction. The reaction mixture is examined directly by IR and does not require sample workup. Data acquisition time is less than one minute. The method has been validated for specificity, precision and accuracy. The results obtained by the FT-IR method for known mixtures and in-process samples compare favorably with those from a normal-phase HPLC method.© (1992) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.
更多
查看译文
关键词
data acquisition,real time
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要