Evaluation Of Differences In Ms/Ms Spectra From Two Different-Generation Q-Tof Instruments: Implications For Metabolomic Database Use

CURRENT PHARMACEUTICAL ANALYSIS(2015)

引用 2|浏览6
暂无评分
摘要
Generating reproducible fragmentation patterns is a problem in electrospray ionization mass spectrometry. Two generations of quadrupole time-of-flight(Q-TOF) instruments were selected to evaluate the reproducibility of tandem mass spectrometry(MS/MS) spectra: an old-and a new-generation instrument. Ninety-four MS/MS and MS/MS/MS spectra with information-rich product ions were obtained in positive-or negative-ion mode with two different-generation Q-TOF instruments, BioTOF-Q and micrOTOF-Q. Two principles were selected to optimize the collision energy. On the one hand, maximal structural information is obtained when the collision energy is increased; on the other hand, when most product ions show relative abundance of more than 10%, the lowest collision energy should be selected. The Paired-Samples T Test method was used to analysis these paired-spectra. We found that most product ions were the same and the profiles of the product ions were similar with both Q-TOF instruments. Interestingly, at the low signal to-noise ratio (S/N) of the precursor ion (ion intensity low to 180 counts), the profile of the MS/MS spectrum is very similar to that at a high S/N. Furthermore, when spectra were generated with micrOTOF-Q even before and after a four-year interval, the profiles and product ions were virtually identical. In summary, databases built with different-generation Q-TOF instruments may be universally applicable.
更多
查看译文
关键词
Reproducibility, two different-generation, quadrupole time-of-flight, tandem mass spectrometry, paired-samples t test, databases built
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要