Novel glass ceramic-type micropattern gas detector with PEG3C

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2013)

Cited 6|Views19
No score
Abstract
A new glass plate (GP) has been developed for a hole-type micropattern gas detector (MPGD). The material of the GP is a crystallized photosensitive etching glass (PEG3C) made by Hoya Corporation. A basic performance test of the hole-type MPGD was carried out using an X-ray beam with Ne-based gas mixtures at 1atm. Gains of up to 5×103 and 6×104 were obtained using a single and a double GP, respectively. An energy resolution of 17% was obtained for 6keV X-rays with a single GP. It was also shown that there is little charging-up effect in the GP. In this paper, we report the characteristics of the new hole-type MPGD with a GP made of PEG3C.
More
Translated text
Key words
Gaseous PMT,Glass ceramic plate,Micropattern gas detector