The Effect of Energy Resolution on the Extraction of Information Content from Gamma-Ray Spectra
Nuclear instruments and methods in physics research Section A, Accelerators, spectrometers, detectors and associated equipment/Nuclear instruments & methods in physics research Section A, Accelerators, spectrometers, detectors and associated equipment(2011)
Abstract
Rapid unambiguous identification of illicit radioactive materials is a matter of international concern that has yet to be fully attained. This is particularly true for shielded fissile materials in the presence of background radiation and routinely encountered benign radionuclides. We present a systematic treatment of the effect of detector resolution on the identification of nuclides in simple and complex shielded gamma-ray spectra. Case studies of some problematic spectra with 1000 counts suggest that, at moderate count rates, near-unambiguous identification (< 1% probability of misidentification) of potential fissile sources requires a detector with an energy resolution of at least 1-2% at 662 key. While somewhat limited in scope, the study establishes a quantitative basis for the comparative evaluation of detectors of comparable efficiency. (C) 2011 Elsevier B.V. All rights reserved.
MoreTranslated text
Key words
Gamma-ray spectrometry,Radionuclide identification,Nuclear security