Analysis of HbA1c on an automated multicapillary zone electrophoresis system.

SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION(2017)

引用 5|浏览21
暂无评分
摘要
Hemoglobin A1c (HbA1c) is a frequently requested laboratory test and there is thus a need for high throughput instruments for this assay. We evaluated a new automated multicapillary zone electrophoresis instrument (Capillarys 3 Tera, Sebia, Lisses, France) for analysis of HbA1c in venous samples. Routine requested HbA1c samples were analyzed immunologically on a Roche c6000 instrument (n=142) and then with the Capillarys 3 Tera instrument. The Capillarys 3 Tera instrument performed approximately 70 HbA1c tests/hour. There was a strong linear correlation between Capillarys 3 Tera and Roche Tina-Quant HbA1c Gen 3 assay (y=1.003x - 0.3246 R-2=.996). The total CV for the 12 capillaries varied between 0.8 and 2.2% and there was a good agreement between duplicate samples (R-2=.997). In conclusion, the Capillarys 3 Tera instrument has a high assay capacity for HbA1c. It has a good precision and agreement with the Roche Tina-Quant HbA1c method and is well suited for high volume testing of HbA1c.