Chrome Extension
WeChat Mini Program
Use on ChatGLM

Amperometric Detector for Gas Chromatography Based on a Silica Sol-Gel Solid Electrolyte

Talanta(2017)

Cited 4|Views20
No score
Abstract
An electrochemical cell comprising a silica sol-gel solid electrolyte, a working electrode that protrudes into a gas phase, and reference and counter electrodes that contact the solid electrolyte comprises an amperometric detector for gas chromatography. Under potentiostatic conditions, a current related to the concentration of an analyte in the gas phase is produced by its oxidation at the three-phase boundary among the sol-gel, working electrode, and the gas phase. The sol-gel is processed to contain an electrolyte that also serves as a humidistat to maintain a constant water activity even in the presence the gas chromatographic mobile phase. Response was demonstrated toward a diverse set of analytes, namely hydrogen, 1,2-ethandithiol, phenol, p -cresol, and thioanisole. Using flow injection amperometry of hydrogen with He as the carrier gas, 90% of the steady-state current was achieved in < 1 s at a flow rate of 20 mL min(-1). A separation of 1,2-ethandithiol, phenol, p -cresol, and thioanisole at a 2.2 mL min(-1) flow rate was achieved with respective detection limits (k = 3 criterion) of 4, 1, 3, and 70 ppmv when the working electrode potential was 800 mV.
More
Translated text
Key words
Amperometric detector,Gas phase,Constant applied potential,Silica sol-gel electrolyte