An automatic approach for artifacts detection and shadow enhancement in intravascular ultrasound images

Signal, Image and Video Processing(2017)

引用 1|浏览21
暂无评分
摘要
Intravascular ultrasound (IVUS) is clinically available for visualizing coronary arteries. However, it suffers from acoustic shadow areas and ring-down artifacts as two of the common issues in IVUS images. This paper introduces an approach which can overcome these limitations. As shadow areas were displayed behind hard plaques in the IVUS grayscale images, calcified plaques were first segmented by using Otsu threshold. Then, active contour, histogram matching, and local histogram matching are implemented. In addition, a new modified circle Hough transform is introduced to remove the ring-down artifacts from IVUS images. In order to evaluate the efficacy of this new method in detection of shadow and ring-down regions, 300 IVUS images are considered. Sensitivity of 89
更多
查看译文
关键词
Intravascular ultrasound (IVUS) images, Otsu thresholding, Shadow region, Active contour, Histogram matching