Validation of the doubly labeled water method using off-axis integrated cavity output spectroscopy and isotope ratio mass spectrometry.

AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM(2018)

引用 21|浏览21
暂无评分
摘要
When the doubly labeled water (DLW) method is used to measure total daily energy expenditure (TDEE), isotope measurements are typically performed using isotope ratio mass spectrometry (IRMS). New technologies, such as off-axis integrated cavity output spectroscopy (OA-ICOS) provide comparable isotopic measurements of standard waters and human urine samples, but the accuracy of carbon dioxide production (<(V) over dot>CO2) determined with OA-ICOS has not been demonstrated. We compared simultaneous measurement <(V) over dot>CO2 obtained using whole-room indirect calorimetry (IC) with DLW-based measurements from IRMS and OA-ICOS. Seventeen subjects (10 female; 22 to 63 yr) were studied for 7 consecutive days in the IC. Subjects consumed a dose of 0.25 g H-2 O-18 (98% APE) and 0.14 g (H2O)-H-2 (99.8% APE) per kilogram of total body water, and urine samples were obtained on days 1 and 8 to measure average daily <(V) over dot>CO2 using OA-ICOS and IRMS. <(V) over dot>CO2 was calculated using both the plateau and intercept methods. There were no differences in <(V) over dot>CO2 measured by OA-ICOS or IRMS compared with IC when the plateau method was used. When the intercept method was used, <(V) over dot>CO2 using OA-ICOS did not differ from IC, but <(V) over dot>CO2 measured using IRMS was significantly lower than IC. Accuracy (similar to 1-5%), precision (similar to 8%), intraclass correlation coefficients (R = 0.87-90), and root mean squared error (30-40 liters/day) of <(V) over dot>CO2 measured by OA-ICOS and IRMS were similar. Both OA-ICOS and IRMS produced measurements of <(V) over dot>CO2 with comparable accuracy and precision compared with IC.
更多
查看译文
关键词
adult,deuterium,humans,oxygen isotope,respiratory gas exchange
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要