NEUTRON WORKPLACE SPECTROMETRY (ENERGY AND DIRECTION) USING TL DETECTORS: FIRST APPROACH AND RESPONSE FUNCTIONS.

D Radeck,M Luszik-Bhadra, T Haninger,M Reginatto

RADIATION PROTECTION DOSIMETRY(2018)

引用 0|浏览16
暂无评分
摘要
The response of albedo dosemeters depends on the energy and angle of the incident neutron radiation. For their use as personal dosemeters, a field-calibration factor has to be applied. The presently used single sphere method for field calibration can be extended and optimised by putting five albedo dosemeters on the surface of a polyethylene sphere and two TL cards in the centre. To investigate the potential of this extension, reference measurements and Monte Carlo calculations were performed and the fluence response of the detectors at different positions on and within the sphere was determined. Calculated response functions demonstrate that information on the energy and directional distribution of neutron fluence can be extracted with this simple set-up for unknown neutron fields.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要