Detection efficiency calculations using Geant4 for a broad-energy germanium gamma spectrometer

Journal of Radioanalytical and Nuclear Chemistry(2017)

引用 5|浏览7
暂无评分
摘要
In assistance of radionuclide measurements at Canada’s Comprehensive Nuclear-Test-Ban Treaty (CTBT) laboratory, a Geant4 Monte Carlo application has been developed in simulating a broad-energy germanium detector and calculating detection efficiencies. The detector model was optimized in a reliable and non-biased manner through simultaneous tuning on gap distance and detector dimension, and was validated over various realistic measurement scenarios. All work is based on a series of experiments which covers the typical energy range of gamma radiation in environmental analysis, and considers the variety of the CTBT sample type, dimension and distance-to-detector. In all cases, the predicted efficiencies are consistent with the empirical ones within 5%, with a typical deviation of 3% in majority.
更多
查看译文
关键词
Broad-Energy Germanium detector,Monte Carlo simulation,Geant4,Efficiency calibration,Comprehensive Nuclear-Test-Ban Treaty