Harmonization Study Between LC-MS/MS and Diasorin RIA for Measurement of 25-Hydroxyvitamin D Concentrations in a Large Population Survey.

JOURNAL OF CLINICAL LABORATORY ANALYSIS(2017)

引用 14|浏览22
暂无评分
摘要
Background: Population-based research on vitamin D has increased dramatically in recent years. Such studies are typically reliant on assay procedures to measure reliable and comparable levels of 25-hydroxyvitamin D [25(OH)D] concentrations. Methods: Concentrations of 25(OH)D-3 and 25(OH)D-2 were measured using LC-MS/MS in 5,915 participants (aged 31 years) of Northern Finland Birth Cohort 1966. Blood samples were assayed in batches over a course of 18 months. As anomalies were present in the measurements, 200 samples were reassayed using Diasorin RIA. Agreement between measurements was assessed by Passing-Bablok regression and limits of agreement (LoA). To harmonize LC-MS/MS with Diasorin RIA measurements, formulae were derived from the LoA. Results: Concentrations measured by LC-MS/MS were much higher than those measured by Diasorin RIA, with a mean difference of 12.9ng/ml. Constant variation was evident between batch measurements after log transformation. Statistical formula was applied separately for each batch of LC-MS/MS measurements, enabling us to remove both the constant and proportional bias that was evident prior to the transformation. Conclusion: Despite the introduction of schemes/programs to improve accuracy of assays to measure 25(OH)D, significant differences can still happen. In these instances, methods to harmonize measurements based on a relatively small number of replicates can be successfully applied to establish confidence and to enable between-study comparisons. (C) 2016 The Authors Journal of Clinical Laboratory Analysis Published by Wiley Periodicals, Inc.
更多
查看译文
关键词
25(OH)D,assay,limits of agreement,population study,reliability replication,vitamin D
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要