Verifying Tumor Position During Stereotactic Body Radiation Therapy Delivery Using (Limited-Arc) Cone Beam Computed Tomography Imaging.

Radiotherapy and oncology(2017)

引用 13|浏览9
暂无评分
摘要
Background and purposeProof of tumor position during stereotactic body radiotherapy (SBRT) delivery is desirable. We investigated if cone-beam CT (CBCT) scans reconstructed from (collimated) fluoroscopic kV images acquired during irradiation could show the dominant tumor position.Materials and methodsFull-arc CBCT scans were reconstructed using FDK filtered back projection from 38kV fluoroscopy datasets (16 patients) continuously acquired during volumetric modulated spine SBRT. CBCT-CT match values were compared to the average spine offset values found using template matching+triangulation of the individual kV images. Multiple limited-arc CBCTs were reconstructed from fluoroscopic images acquired during lung SBRT of an anthropomorphic thorax phantom using 20–180° arc lengths and for 3 breath-hold lung SBRT patients.ResultsDifferences between 3D CBCT-CT match results and average spine offsets found using template matching+triangulation were 0.1±0.1mm for all directions (range: 0.0–0.5mm). For limited-arc CBCTs of the thorax phantom, the automatic 3D CBCT-CT match results for arc lengths of 80–180° were ≤1mm. 20° CBCT reconstruction still allowed for positional verification in 2D.Conclusions(Limited-arc) CBCT reconstructions of kV images acquired during irradiation can identify the dominant position of the tumor during treatment delivery.
更多
查看译文
关键词
Cone beam computed tomography,Fluoroscopy,Stereotactic body radiation therapy,Intrafraction monitoring,Positional verification
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要