Fallacies and Pitfalls of Dialysis Sodium Prescription and Control.

BLOOD PURIFICATION(2018)

引用 14|浏览11
暂无评分
摘要
Dialysate conductivity is routinely used as a surrogate for dialysate sodium concentration. However, dialysis machine manufacturers apply different conductivity temperature correction coefficients. With the same conductivity in dialysis machines manufactured by different manufacturers, dialysate sodium may significantly differ. Also, electrolyte prescriptions are individualized (K, Ca, HCO3) and this is associated with another variation in dialysate sodium in the order of 1-5 mmol/L and both deviations are cumulative and chronic for each patient. Equivalence of the prescribed dialysate sodium and the concentration measured in it is not granted. Both variables differ and it is machine dependent! This paper analyses those variations from a technical point of view and suggests how to detect them and how to deal with or avoid them in clinical practice. (C) 2018 S. Karger AG, Basel
更多
查看译文
关键词
Dialysis sodium prescription,Dialysate sodium,Dialysate conductivity,Dialysate sodium correction,Conductivity cell monitoring