Multicenter investigation of the incidence of inferior vena cava filter fracture

Japanese journal of radiology(2018)

引用 6|浏览25
暂无评分
摘要
Purpose Inferior vena cava filter fracture (FF) may cause life-threatening complications, including cardiac tamponade, although the actual prevalence remains unclear. Therefore, we investigated the incidence of FF. Materials and methods Data on fracture incidence with filter brands, filter positions [suprarenal (SR) vs. infrarenal (IR)], and follow-up durations were collected from the databases of eight hospitals. Results Of 532 patients, Günther Tulip (GT), Trap/OptEase (TE/OE), ALN and VenaTech (VT) were implanted in 345, 147, 38 and 2 patients, respectively. Of these, filter retrieval was attempted in 110 (21.7%) patients and was successful in 106 (96.4%). Of the remaining 426 patients, FFs were observed in two (0.7%) of 270 GT filters and 19 (14.1%) of 135 TE/OE filters. Fragment embolization occurred in one patient with a GT filter (50.0%) and three with a TE/OE filter (15.8%) with a total follow-up interval of 718.0 ± 1019.4 days. FF occurred more frequently in TE/OE than in GT filters ( p < 0.001). Kaplan–Meier estimates showed significantly higher fracture-free rates for GT than TE/OE ( p < 0.001) and IR-TE/OE than SR-TE/OE ( p < 0.05). Conclusions TE/OE filters are not suitable for permanent implantation due to the relatively early and high fracture rates.
更多
查看译文
关键词
Deep vein thrombosis,Fracture,Inferior vena cava filter,Migration,Pulmonary embolism
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要