Resolution and registration in dual-plane co-RASOR MR.
PHYSICS IN MEDICINE AND BIOLOGY(2018)
摘要
Magnetic resonance imaging (MRI) has superior soft tissue contrast and lower interobserver variability compared to computed tomography and advances in equipment and pseudo-CT estimation have allowed for MR-only radiation therapy planning. Dedicated MR sequences have been used to localize paramagnetic structures with positive contrast, and most implanted seeds are gold fiducial markers (GFMs). We used a fast, dual-plane co-RASOR sequence to localize implanted GFMs with positive contrast in phantom and tissue to assess their resolution and registration accuracy of registration to CT. Off-resonant reconstructions of co-RASOR images were able to resolve GFMs down to 5 mm apart at 12 cm FOV. No systematic biases were observed by comparing registration of co-RASOR and bSSFP to CT images in an MR-compatible Lego phantom with a set of highly visible known points. The standard deviations of the MR to CT distance errors were <0.5 mm in all directions. We separated the component due to registration by comparing the two MR sequences, which had a maximum standard deviation of 0.36 mm in the SI-direction. Registration using the positive contrast points in a porcine sample phantom showed increased errors, but co-RASOR still performs acceptably with a target registration error of <0.75 mm. The dual-plane co-RASOR sequence could then be used for both registration and image tracking when performing MR-only radiation therapy planning.
更多查看译文
关键词
magnetic resonance imaging (MRI),gold fiducial marker,co registration,computed tomography (CT),co RASOR