Initial Beam Emission Spectroscopy Diagnostic System On Hl-2a Tokamak

REVIEW OF SCIENTIFIC INSTRUMENTS(2018)

引用 19|浏览101
暂无评分
摘要
A beam emission spectroscopy system is being developed and deployed on the HL-2A tokamak to measure local low wavenumber (k(perpendicular to)rho(i) < 1) density fluctuations by measuring the Doppler-shifted emission from a 50 kV deuterium heating neutral beam. High spatial resolution (Delta r <= 1 cm, Delta z <= 1.5 cm) measurements are achieved with customized in-vacuum optics. High frequency, high-gain preamplifiers sample the light intensity at a Nyquist frequency of 1 MHz and achieve a high S/N ratio via high optical throughput, low-noise preamplifiers, and high quantum efficiency photodiodes. A first set of 16 detector channels [configured in an 8 (radial) x 2 (poloidal) array] has been installed and tested at HL-2A, covering the radial range r/a = 0.8-1.1. The frequency and wavenumber spectra have been measured under different plasma conditions. Initial measurements have demonstrated the capability of measuring edge plasma density fluctuation spectra and the poloidal flow velocity fields with a high S/N ratio. Published by AIP Publishing.