Applicability of refractometry for fast routine checking of hospital preparations.

European Journal of Pharmaceutical Sciences(2016)

引用 2|浏览4
暂无评分
摘要
Quality control of hospital pharmacy formulations is of the utmost importance to ensure constant quality and to avoid potential mistakes before administration to the patient. In this study we investigated the applicability of refractometry as a fast, inexpensive and easy-to-use quality control measurement. Refractive indices (RI) of a multitude of different hospital formulations with varying concentrations of active compound were measured. The samples consisted of a number of binary aqueous solutions (one compound in water), complex aqueous solutions (multiple compounds in water or in a constant matrix), two suspensions and one emulsion. For all these formulations, linear regression analysis was performed, quality control limits determined and accuracy and repeatability were checked. Subsequently, actual hospital pharmacy samples were analyzed to check whether they were within the specified limits. For both binary and complex aqueous formulations, repeatability was good and a linear correlation for all samples could be observed on condition that the concentration of the active compound was sufficiently high. The refractometer was not sensitive enough for solutions of folic acid and levothyroxine, which had too low a concentration of active compound. Due to lack of homogeneity and light scattering, emulsions and suspensions do not seem suitable for quality control by refractometry.
更多
查看译文
关键词
Refractometry,Patient drug formulations,Quality control,Hospital pharmacy