Technical Note: Magnetic field effects on Gafchromic-film response in MR-IGRT.

Medical physics(2016)

引用 36|浏览12
暂无评分
摘要
The presence of a magnetic field affects crystal orientation and polymerization during irradiation, where netOD decreased by an average of 8.7%, 8.0%, and 4.3% in the red, green, and blue channels, respectively. The under response was dependent on dose and differed by up to 15% at 17.6 Gy.
更多
查看译文
关键词
MR-IGRT,ViewRay,Gafchromic,EBT2,Radiochromic