Light field otoscope design for 3D in vivo imaging of the middle ear.

BIOMEDICAL OPTICS EXPRESS(2017)

引用 42|浏览23
暂无评分
摘要
We present a light field digital otoscope designed to measure three-dimensional shape of the tympanic membrane. This paper describes the optical and anatomical considerations we used to develop the prototype, along with the simulation and experimental measurements of vignetting, field curvature, and lateral resolution. Using an experimental evaluation procedure, we have determined depth accuracy and depth precision of our system to be 0.05-0.07 mm and 0.21-0.44 mm, respectively. To demonstrate the application of our light field otoscope, we present the first three-dimensional reconstructions of tympanic membranes in normal and otitis media conditions, acquired from children who participated in a feasibility study at the Children's Hospital of Pittsburgh of the University of Pittsburgh Medical Center. (C) 2016 Optical Society of America