Vacuum arc plasma deposition of thin titanium dioxide films on silicone elastomer as a functional coating for medical applications.

Cécile Boudot, Marvin Kühn, Marina Kühn-Kauffeldt,Jochen Schein

Materials Science and Engineering: C(2017)

引用 22|浏览1
暂无评分
摘要
Silicone elastomer is a promising material for medical applications and is widely used for implants with blood and tissue contact. However, its strong hydrophobicity limits adhesion of tissue cells to silicone surfaces, which can impair the healing process. To improve the biological properties of silicone, a triggerless pulsed vacuum cathodic arc plasma deposition technique was applied to deposit titanium dioxide (TiO2) films onto the surface. Scanning electron microscopy, atomic force microscopy, X-ray photoelectron spectroscopy, Raman spectroscopy and contact angle measurements were used for coating characterization. Deposited films were about 150nm thick and exhibited good adhesion to the underlying silicone substrate. Surface wettability and roughness both increased after deposition of the TiO2 layer. In addition, cell-biological investigations demonstrated that the in-vitro cytocompatibility of TiO2-coated samples was greatly improved without impacting silicone's nontoxicity. For validation of use in medical devices, further investigations were conducted and demonstrated stability of surface properties in an aqueous environment for a period of 68days and the coating's resistance to several sterilization methods.
更多
查看译文
关键词
Silicone elastomer,Titanium dioxide,Coating,Vacuum arc plasma,Cytocompatibility,Cytotoxicity