Fluorescence microscope light source stability

Firas Mubaid,Daniel Kaufman,Tse-Luen Wee, Dong-Son Nguyen-Huu, David Young, Maria Anghelopoulou,Claire M. Brown

Histochemistry and Cell Biology(2019)

引用 8|浏览1
暂无评分
摘要
The process of fluorescence starts with the efficient generation of light that is required for the excitation of fluorophores. As such, light sources are a crucial component of a fluorescence microscope. Choosing the right illumination tool can not only improve the quality of experimental results, but also the microscope’s economic and environmental footprint. While arc lamps have historically proven to be a reliable light source for widefield fluorescence microscopy, solid-state light-emitting diodes (LEDs) have become the light source of choice for new fluorescence microscopy systems. In this paper, we demonstrate that LEDs have superior light stability on all timescales tested and use less electrical power than traditional light sources when used at lower power outputs. They can be readily switched on and off electronically, have a longer lifetime and they do not contain mercury, and thus are better for the environment. We demonstrate that it is important to measure light source power output during warm-up and switching, as a light source’s responsiveness (in terms of power) can be quite variable. Several general protocols for testing light source stability are presented. A detailed life cycle analysis shows that an LED light source can have a fourfold lower environmental impact when compared to a metal halide source.
更多
查看译文
关键词
Stability,Light source,Solid state,Fluorescence,Microscopy,LED