The individual systematic difference between CoaguChek and STA-SPA.

Arne Åsberg, Hilde Hegseth, Maria Averina,Lena Løfblad

SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION(2019)

引用 4|浏览4
暂无评分
摘要
Using CoaguChek to measure PT-INR and comparing the results with those from the hospital laboratory, some patients get consistent results while others do not. The extent of this problem is unknown. Our study aimed to quantify the between-subject variation of the systematic PT-INR difference between CoaguChek and a hospital laboratory method. We used register data with PT-INR results from both CoaguChek and a hospital laboratory method (STA-SPA+) in samples taken simultaneously from 108 patients. After excluding five patients with outlying results, we used mixed-effects models to estimate individual slopes and intercepts to describe the systematic relationship between the two methods for each patient, and calculated the fraction of patients having a systematic difference greater than 0.3 INR units. The included 103 patients had from 3 to 16, median seven data pairs measured over a time span from 15 to 2319, median 234 days. The mean of individual slopes was 1.113, with a standard deviation of 0.137. Corresponding values for the intercept were -0.151 and 0.208, respectively. Adjusted for the average systematic difference, the proportion of patients with a systematic difference greater than 0.3 INR units increased from 15% at a PT-INR level of 2.5 to 50% at a PT-INR level of 4. The systematic difference between CoaguChek and STA-SPA + varies considerably between patients. This precludes using a single, common formula to make the CoaguChek results directly comparable to the results from the hospital laboratory.
更多
查看译文
关键词
Between-subject variation,CoaguChek,international normalized ratio,method comparison,point-of-care testing,warfarin
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要