Volumetric synthetic aperture imaging with a piezoelectric 2D row-column probe

Proceedings of SPIE(2016)

引用 3|浏览11
暂无评分
摘要
The synthetic aperture (SA) technique can be used for achieving real-time volumetric ultrasound imaging using 2-D row-column addressed transducers. This paper investigates SA volumetric imaging performance of an in-house prototyped 3 MHz lambda/2-pitch 62+62 element piezoelectric 2-D row-column addressed transducer array. Utilizing single element transmit events, a volume rate of 90 Hz down to 14 cm deep is achieved. Data are obtained using the experimental ultrasound scanner SARUS with a 70 MHz sampling frequency and beamformed using a delay-and-sum (DAS) approach. A signal-to-noise ratio of up to 32 dB is measured on the beamformed images of a tissue mimicking phantom with attenuation of 0.5 dB cm(-1) MHz(-1), from the surface of the probe to the penetration depth of 300 lambda. Measured lateral resolution as Full-Width-at-Half-Maximum (FWHM) is between 4 lambda and 10 lambda for 18 % to 65 % of the penetration depth from the surface of the probe. The averaged contrast is 13 dB for the same range. The imaging performance assessment results may represent a reference guide for possible applications of such an array in different medical fields.
更多
查看译文
关键词
3-D ultrasound imaging,2-D row-column addressed transducer,synthetic aperture (SA)