Guidelines for reducing image retakes of general digital radiography

ADVANCES IN MECHANICAL ENGINEERING(2016)

引用 19|浏览9
暂无评分
摘要
Image retake of radiological examinations not only increases the risk of radiation exposure of the patients, but also wastes the medical resource and degrades the quality of services of the hospitals. This study aimed at discovering factors affecting image retake of general digital radiography for setting guidelines to reduce the image retaking rate. A total of 98,503 general X-ray images retrieved from the picture archiving and communication system database of a medical center in central Taiwan were analyzed. The results showed that the total retaking rate was 4.89% with the position error (56.05%) was the main factor causing image retakes and chest examination showed the highest frequency (1544 cases). On the other hand, skull/face exhibited the highest retaking rate (9.81%) among various types of examinations. After discovering the factors affecting the image retaking rate, suitable guidelines were proposed and introduced. The image retake rate had been significantly reduced to 4.38% and 3.57% 1month and 6months, respectively, after the introduction of guidelines. In conclusion, image retake analysis is a quality indicator and is effective for quality assurance of digital radiology. Regular analysis of image retake can find factors inducing image retake and is useful for designing guidelines to reduce the image retake rate.
更多
查看译文
关键词
General digital radiography,image retake,quality assurance