High Performance Gamma-Ray Spectrometer for Runaway Electron Studies on the Ft-2 Tokamak

Nuclear instruments and methods in physics research Section A, Accelerators, spectrometers, detectors and associated equipment/Nuclear instruments & methods in physics research Section A, Accelerators, spectrometers, detectors and associated equipment(2016)

Cited 28|Views13
No score
Abstract
A gamma-ray spectrometer based on LaBr3(Ce) scintillator has been used for measurements of hard X-ray emission generated by runaway electrons in the FT-2 tokamak plasmas. Using of the fast LaBr3(Ce) has allowed extending count rate range of the spectrometer by a factor of 10. A developed digital processing algorithm of the detector signal recorded with a digitizer sampling rate of 250MHz has provided a pulse height analysis at count rates up to 107s−1. A spectrum deconvolution code DeGaSum has been applied for inferring the energy distribution of runaway electrons escaping from the plasma and interacting with materials of the FT-2 limiter in the vacuum chamber. The developed digital signal processing technique for LaBr3(Ce) spectrometer has allowed studying the evolution of runaways energy distribution in the FT-2 plasma discharges with time resolution of 1–5ms.
More
Translated text
Key words
LaBr3(Ce)detector,Hard X-ray spectrometry,Tokamak,Diagnostics of runaway electrons,Physics of plasma,Hard X-ray spectrum deconvolution