A New Hemostatic Device Utilizing A Novel Transmission Structure For Delivery Of Adrenaline And Microwave Energy At 5.8 Ghz

INTERNATIONAL JOURNAL OF MICROWAVE AND WIRELESS TECHNOLOGIES(2017)

引用 4|浏览8
暂无评分
摘要
A novel transmission line structure has been developed to facilitate the delivery of both adrenaline and microwave energy to achieve hemostasis. A proximal end impedance transformer and radiative tip have been designed and manufactured to provide good match between the novel hollow transmission line and the microwave source and tissue, respectively. Further consideration of the challenges and problems encountered along with evidence of successful microwave energy delivery at 5.8 GHz into porcine liver model providing a controlled and focused coagulation zone of approximately 5 mm.
更多
查看译文
关键词
Biomedical applications, Antenna design, Modeling and measurements