Multi-parameter calibration of streak/framing camera based on integrated optical system with delay and attenuation modules

Optik(2017)

引用 3|浏览0
暂无评分
摘要
We demonstrate a multi-parameter calibration method of the streak camera and the framing camera with the help of beam division delay and attenuation devices in an integrated optical system based on sub-picosecond short-pulse UV laser. In the designed optical scheme, the four operating modes employed the optical delay, beam division delay and attenuation, and etalon modules can be switched to synchronize the UV laser pulse with the trigger signal of diagnosis instruments, calibrate the temporal resolution and dynamic range of framing cameras, and correct the sweep speed of streak cameras, respectively. The results show that the delay time error of ±0.1ns in the optical delay module, the uniform distribution of sub-beams with the delay time precision of ±0.03ps and isocon descending energy in the beam division delay and attenuation modules, and pulse broadening of ∼22% in the whole system are achieved. In the meantime, the experiment results display the sweep speed of 1.2ps/pixel in a streak camera and the dynamic range of 131 in a framing camera. The proposed folding optical system with high-precision, low-loss and multi-function can be extensively applied in the temporal calibration of diagnosis instruments such as streak cameras and framing cameras.
更多
查看译文
关键词
Optical design,Calibration,Sweep speed,Temporal resolution,Dynamic range,Camera
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要