Calibration System For Low-Current Transresistance Amplifiers With A Low-Frequency Sinewave Source

2017 XXXIIND GENERAL ASSEMBLY AND SCIENTIFIC SYMPOSIUM OF THE INTERNATIONAL UNION OF RADIO SCIENCE (URSI GASS)(2017)

引用 0|浏览2
暂无评分
摘要
The calibration of the transresistance gain of low-current dc amplifiers can be performed with the capacitance-charging method, where the calibration current is generated by a calibrated voltage ramp applied to a reference capacitor. The method is typically implemented with a purposely-built voltage source having a very low-frequency trapezoidal periodic waveform output. Here we propose a variation of the method based on a sinusoidal excitation, that can be provided by a commercial oscillator and therefore more easily implemented. Sampled sinusoidal input and output are processed with a seven-parameter sine-fitting algorithm. The comparison between calibration with sinusoidal and trapezoidal excitations is ongoing and will be presented at the Conference.
更多
查看译文
关键词
trapezoidal excitations,sinusoidal excitations,sinusoidal excitation,low-frequency trapezoidal periodic waveform output,voltage source,reference capacitor,calibrated voltage ramp,calibration current,capacitance-charging method,dc amplifiers,transresistance gain,low-frequency sinewave source,low-current transresistance amplifiers,calibration system