谷歌浏览器插件
订阅小程序
在清言上使用

Fading time and irradiation dose estimation from thermoluminescent dosemeters using glow curve deconvolution

Radiation Measurements(2018)

引用 16|浏览1
暂无评分
摘要
Thermoluminescence dosemeters are commonly used in various fields of dosimetry. During the elapsed time between irradiation and readout, dosemeters are affected by thermal fading. As a result the measured signal is reduced resulting in an underestimation of the irradiation dose. This is a challenging issue, especially in personal dosimetry. Different techniques have been developed in the past to overcome the influence of fading. These techniques often come along with a loss of information which reduces the accuracy of the irradiation dose estimation. In this work, a method is developed which is based on glow curve deconvolution and which results in an irradiation dose estimation with no fading time dependence. The method also gives an estimation of the fading time.
更多
查看译文
关键词
TL-DOS,Thermoluminescence,Glow curve deconvolution,Fading time estimation,Irradiation dose estimation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要