谷歌浏览器插件
订阅小程序
在清言上使用

Effect of High Intensity Focused Ultrasound Transducer F-number and Waveform Non-Linearity on Inertial Cavitation Activity

˜The œJournal of the Acoustical Society of America/˜The œjournal of the Acoustical Society of America(2018)

引用 1|浏览22
暂无评分
摘要
Enhanced chemotherapeutic drug delivery has been shown using pulsed high intensity focused ultrasound (pHIFU) without contrast agents. The threshold of these inertial cavitation effects has recently been correlated to the formation of shocks at the focus. The shock amplitude and corresponding peak negative pressure (p-) are primarily determined by the transducers F-number with less focused (higher F-number) transducers producing shocks at lower p-. Here, the dependence of inertial cavitation activity on F-number was investigated in gel phantoms using passive cavitation detection (PCD) and high-speed photography. HIFU transducers with the same aperture but different F-numbers (0.77, 1.02, and 1.52), operable at three driving frequencies (1 MHz, 1.5 MHz, and 1.9 MHz), were utilized with driving conditions consisting of 1 ms pulses delivered every second and p- from 1 to 15 MPa. Broadband noise emissions recorded by PCD were batch-processed to extract cavitation probability and persistence while concurrent imaging was performed in the focal pHIFU region. At the same p-, both PCD metrics and the imaging revealed enhanced cavitation activity at higher F-numbers. These results support the use of less focused, smaller-footprint transducers for achieving desired cavitation-aided drug delivery. [Work supported by NIH R01EB023910, K01DK104854, R01EB007643, and RFBR 17-54-33034.]
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要