A Liquid Scintillation Detector for Radioassay of Gadolinium-Loaded Liquid Scintillator for the LZ Outer Detector

Nuclear instruments and methods in physics research Section A, Accelerators, spectrometers, detectors and associated equipment/Nuclear instruments & methods in physics research Section A, Accelerators, spectrometers, detectors and associated equipment(2019)

引用 11|浏览7
暂无评分
摘要
We report on the design and performance of the LUX-ZEPLIN (LZ) "Screener", a small liquid scintillator detector consisting of approximate to 23 kg of LAB-based gadolinium-loaded liquid scintillator (GdLS) to be used in the LZ Outer Detector. The Outer Detector will be filled with 17.3 tonnes of GdLS and will surround the central liquid xenon time projection chamber of LZ. Its primary function will be to tag neutron events in the liquid xenon which could mimic a WIMP dark matter signal. To meet the deadtime requirements for the Outer Detector, the radioimpurity levels in the GdLS must be kept below less than or similar to 0.07 mBq/kg. This background level corresponds to a rate of approximate to 50 Hz above an energy threshold of 100 keV. The Screener was operated in the ultra-low-background environment of the former LUX water shield in the Davis Laboratory at the Sanford Underground Research Facility for radioassay of the GdLS. Careful selection of detector materials and use of ultra-low-background PMTs allows the measurement of a variety of radioimpurities. The C-14/C-12 ratio in the scintillator is measured to be (2.83 +/- 0.06(stat.)+/- 0.01(sys.)) x 10(-1)7. Use of pulse shape discrimination allows the concentration of isotopes throughout the U-238, U-235, and Th-232 chains to be measured by fitting the collected spectra from alpha and beta events. We find that equilibrium is broken in the U-238 and Th-232 chains and that a significant portion of the contamination in the GdLS results from decays in the Ac-227 subchain of the U-235 series. Predictions for the singles rate in the Outer Detector are presented. The rate from radioimpurities above 100 keV in the GdLS is estimated to be 97.9 +/- 6.4 Hz, with 65.5 +/- 1.9 Hz resulting from alpha-decays.
更多
查看译文
关键词
Low-background,Gadolinium,Liquid scintillator
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要