Cardiac computed tomography angiography (CCTA) is the gold standard for accurately sizing the aortic valve annulus prior to aortic valve replacement. A reduction of contrast volume administered for CCTA, without sacrificing image quality, is desirable. Signal-to-noise ratio represents final CCTA image quality. Consecutive patients referred to CCTA for aortic valve annulus sizing were retrospectively analyzed. Patients were grouped into a low-dose contrast (LDCT) group and traditional dose contrast (TDCT) group. In the LDCT group, contrast dose was <50% of the maximal allowable dose (3.7 × estimated glomerular filtration rate). Guided by a time-density curve, the contrast was administered in a two-stage infusion, and retrospectively gated images were acquired with a 64-multidetector computed tomography scanner. Out of 123 patients (age 80 ± 9 years; 46% female), 65 (52.9%) underwent LDCT and 58 (47.2%) underwent TDCT. Contrast volume was significantly lower in the LDCT group (LDCT 41.2 ± 9.8 vs TDCT 76.2 ± 14.2 mL; P < 0.001). The signal-to-noise ratio of the aortic root was 10.4 ± 4.1 for the LDCT group and 8.4 ± 3.3 for the TDCT group (P = 0.004). Aortic root dimensions could be measured in both LDCT and TDCT groups. In conclusion, LDCT with 64-slice CCTA can effectively size the aortic valve annulus to direct aortic valve replacement while offering reduced contrast exposure.