Comparison of benchtop pressure gradient measurements in 3D printed patient specific cardiac phantoms with CT-FFR and computational fluid dynamic simulations.

Proceedings of SPIE(2019)

引用 2|浏览11
暂无评分
摘要
Purpose: Various CT-FFR methods are being proposed as a non-invasive method to estimate cardiac disease severity. 3D printed patient specific cardiovascular models with high geometric accuracy can be used to simulate blood flow conditions and perform precise and repeatable benchtop flow experiments for validation of such methods. Materials and Methods: Twelve patient-specific 3D printed cardiac phantoms were created from CT Angiography (CTA) scans using a compliant 3D printing material. Pressure sensors were connected to the aortic root and distal ends of the three main coronary arteries to measure benchtop pressure gradients for each stenosed vessel. The patient geometries were used in Canon Medical Systems 1D fluid dynamics algorithm to calculate the CT-FFR. In addition, a 3D computational fluid dynamics simulation was done using ANSYS to estimate pressure gradients across the coronary arteries. Experimental data and 1D and 3D flow simulations were compared to the standard catheter lab FFR measurement (Invasive-FFR). Results: The average percent difference in Benchtop FFR/Invasive FFR, CT-FFR/Invasive FFR, and CFD-based validation/Invasive FFR was 0.05, 0.06, and 0.07 respectively. The average time it took for the CT-FFR simulation was similar to 35 minutes and it took similar to 15 hours for the CFD-based simulation but can vary based on the number of iterations the user defines the software to run. Conclusions: Benchtop FFR proved to be highly accurate when compared to both 1D and 3D CFD software and therefore, 3D printing of patient specific coronary phantoms is a quality tool for CT-FFR software validation.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要