Geant4 analysis and optimization of a double crystal phoswich detector for beta–gamma coincidence detection

Nuclear Science and Techniques(2018)

引用 2|浏览19
暂无评分
摘要
In this study, a novel phoswich detector for beta–gamma coincidence detection is designed. Unlike the triple crystal phoswich detector designed by researchers at the University of Missouri, Columbia, this phoswich detector is of the semi-well type, so it has a higher detection efficiency. The detector consists of BC-400 and NaI:Tl with decay time constants of 2.4 and 230 ns, respectively. The BC-400 scintillator detects beta particles, and the NaI:Tl cell is used for gamma detection. Geant4 simulations of this phoswich detector find that a 2-mm-thick BC-400 scintillator can absorb nearly all of the beta particles whose energies are below 700 keV. Further, for a 2.00-cm-thick NaI:Tl crystal, the gamma source peak efficiency for photons ranges from a maximum of nearly 90% at 30 keV to 10% at 1 MeV. The self-absorption effect is also discussed in this paper in order to determine the carrier gas’s influence.
更多
查看译文
关键词
Geant4, Phoswich detector, Beta–gamma coincidence detection, Detection efficiency