Cross-link is a risk factor for rod fracture at pedicle subtraction osteotomy site: A finite element study.

Journal of Clinical Neuroscience(2019)

引用 9|浏览4
暂无评分
摘要
•We aim to investigate whether cross-link is a risk factor for rod fracture at the pedicle subtraction osteotomy (PSO) site by using finite element model (FEM).•The geometry and dimensions of FEM were obtained from a high-resolution computed tomography scan data of a 65 years old female patient with spinal sagittal imbalance.•After PSO surgery, PVMS of the rod was significantly increased during flexion and extension at PSO site.•With cross-link at PSO site, PVMS of rod was further significantly increased during flexion and extension.
更多
查看译文
关键词
Cross-link,Rod fracture,PSO,Spine,Finite element analysis